Performance Data
Model ; 3021 3021-ST 3022
Specifical Sensitivity [㎛] 8 ... 12 ㎛ 8 ... 12 ㎛ 2 ... 5 ㎛
Temperature Measuring Range[℃] -40 ... +1200℃ -40 ... +1200℃ -10 ... +1200℃
Thermal Resolution ±0.03℃ ±0.03℃ ±0.12℃
(Black body radiator with 30℃)
Absolute Temperature Measuring ±2K (measurement range up to 100℃)otherwise +1% of range full scale
accuracy[K] at (22±2)℃
ambient temperature
Geometrical Resolution [mrad]
(for standard lens) 1.5 mrad 1.5 mrad 3.0 mrad
Detector Type HgCdTe HgCdTe HgCdTe
Detector Cooling Liquid Nitrogen Stirling Cooler 3 stage thermoelectric
Dimensions(L x B x H)[mm] 240 x 192 x 200 mm 240 x 192 x 200 mm 240 x 192 x 200 mm
Weight [Kg] 5.1 Kg 5.0 Kg 4.7 Kg
Common system features.
Object field (standard lens) 30˚(H) x 20˚(V)
Object field (wide-angle attachment) 60˚(H) x 40˚(V)
Object distance (standard lens) 0.2m ... ∞
Object distance (standard lens) 2m ... ∞
Electro-optical zoom 6:1
Pixels per Line 360
Lines per Frame 240 (100, 50 optional)
Frame repeat ctcle [s] 0.9
Power supply Power pack, internal/external error
A/D-dynamic Range 16 bit
Display 4" TFT (Activ)
Storage Medium External memory card type II, ATA-compatible, internal Flash memory
Video Output RGB-PAL, RGB-NTSC switchable, TV-connection via video cable DSUB->SCART cable
Functions Multipixel Measurements(max. 10 test point),
Frame accumulation (max. 10 frames)
Online minimum/maximum sport
Display functions, switchable 3 Color palettes (256, 32, 16), 2 black-white range (standard/inverse)
isotherm superimposition (max. 9), editable emission value, comment
Automatic Functions Autofocus, autolevel, autorange, autosave
Range of operating temperature [℃] -10 ..... +40℃
Range of storage temperature (unpacked) -25 ..... +70℃